Size & Dimension
- Size: 164 x 163 x 167 mm
- Weight: 0.58Kg
Operating Temp Range
- 17 – 37°C
Power Supply
- From USB port and 9V battery for hygiene fan
Flow Detection Principle
- Fleisch type pneumotachograph
Volume Detection
- Flow integration @ 100Hz
Accuracy
- Volumes: Better than +/- 3%
- Flows: Better than +/- 5% (Max 16L/s / Min 0.02L/s)
Linearity
- +/- 1% in range
- 0.1 L/s to 16 L/s
Computer Requirements: Minimal Specifications
- CPU: Intel Core i5 (equivalent or better)
- RAM: 6 GB or greater
- STORAGE: 1GB for installation, plus 1GB per year for data growth
- Operating System: Windows® XP Professional, 7 or 8 (32/64-bit)
- USB: One available USB port (2.0 or better)
Test capability
Static Spirometry, Flow Volume Loops, MVV, Cough Peak Flow and Bronchial Challenge.
Nonin WristOx
Six Minute Walk
Predicteds
GLI, ATS/ERS, Polgar, Knudson, Crapo & Morris, Hsu, Wang & Dockery and many more.
Report Library
Brilliant library of report options:
- Logo and header design
- Manual entry information
- Numerous test graphics
- Summary of all efforts
- Overlay of graphics
- Past test result history
- Serial data trend graphing
- Computer impressions
- Captured digital signature
- Administrative Reports
- QA Reports
- Research Query Reports
Interfacing
Health Level Seven (HL7)
Remote Interpretation
When tests are completed, results can be routed to any network/internet connected station for physician interpretation
Remote Help
Using secure internet connections